Utilization of anterior segment optical coherence tomography (AS-OCT) enhanced high resolution corneal (EHRC) software in measuring Pterygium thickness
DOI:
https://doi.org/10.31436/imjm.v15i1.1380Abstract
Introduction: To propose an objective method of quantifying pterygium thickness using AS-OCT. Methods: Anterior segment imaging was performed using Enhanced High Resolution Corneal (EHRC) of Visante™ OCT in 68 primary pterygium eyes. Prior to imaging, corneal topography assessment was performed on each pterygium eye in order to identify its topographic location. Based on topography mapping, three meridians (in degrees) were selected as close as possible to the pterygium border, which signify the demarcation of pterygium from the cornea. Reliability testing between intra and inter-observer on OCT imaging modality was examined using Intraclass correlation (ICC) and scatter plot. Results: The overall (n = 68) mean and SD of pterygium thickness via OCT modality were 0.48 ± 0.10 mm (CI: 0.45 – 0.50). OCT imaging also showed excellent intra and intergrader reliability in measuring pterygium thickness with ICC of 0.997 (CI: 0.994 – 0.998). Conclusions: We have described and demonstrated the use of EHRC as an objective measurement for pterygium thickness. This tool proved could eliminate the variability of human inter-graders in assessing thickness of pterygium with its promising ICC. This study had proved that anterior segment OCT imaging is a better choice in assessing pterygium compared to traditional slit-lamp biomicroscopy. We suggest that this tool is applicable for future work such as to understand the role thickness in pterygium morphology and its progression. Better understanding of pterygium morphology is important for future clinical decision, especially on prediction of induced corneal astigmatism and visual impairment due to pterygium.
Downloads
Downloads
Published
How to Cite
Issue
Section
License
All material submitted for publication is assumed to be submitted exclusively to the IIUM Medical Journal Malaysia (IMJM) unless the contrary is stated. Manuscript decisions are based on a double-blinded peer review process. The Editor retains the right to determine the style and if necessary, edit and shorten any material accepted for publication.
IMJM retain copyright to all the articles published in the journal. All final ‘proof’ submissions must be accompanied by a completed Copyright Assignment Form, duly signed by all authors. The author(s) or copyright owner(s) irrevocably grant(s) to any third party, in advance and in perpetuity, the right to use, reproduce or disseminate the research article in its entirety or in part, in any format or medium, provided that no substantive errors are introduced in the process, proper attribution of authorship and correct citation details are given, and that the bibliographic details are not changed. If the article is reproduced or disseminated in part, this must be clearly and unequivocally indicated.